Using the PC application

This page walks through a session with the control application, from launch to the result files. It assumes the application is installed (Installing the PC application) and the box is powered and connected, with the main module in experiment (V2 emulation) mode (Deploying the firmware).

About the screenshots

The screenshots on this page were taken with a simulated box (no hardware connected), so the sensor indicators are all off and the numbers in the stage panel were filled in for illustration. They show the real layout of the current interface.

Your session in six steps

  1. Launch the application (beatbox).

  2. Choose the stage and enter the animal’s details in the first dialog.

  3. Connect to the box: choose the serial port, click Connect.

  4. Choose the experiment folder where results will be saved.

  5. Start the experiment.

  6. Monitor; stop when done (or let the stage finish).

1–2. The start dialog

On launch, a small dialog Select Experiment Stage appears.

The start dialog, filled in for animal M012 at stage S3

Field

What to enter

Experiment stage

S1, S2, S3, S3.1, S4, or All

Animal name

The animal’s identifier. It is written in the file names and in the logs.

Weight (g)

The animal’s weight

Box

The box identifier

Sex

X (unset), Male or Female

Light cycle

X (unset), Red or White

  • Choose a single stage to run just that stage.

  • Choose All to start at S1; when a stage is complete, the application moves automatically to the next one (S1 → S2 → S3 → S3.1 → S4) and shows Experiment complete after S4. See Training stages S1–S4 for what each stage is.

Use identifiers that are unique and consistent across animals and boxes: they end up in your data.

3. Connect to the box

The main window has one panel per box (one by default; several if you started with --n-devices). Before you connect, it looks like this (stage S1):

The main window just after launch, not yet connected

In a panel:

  • Serial Connection: pick the Port (click Refresh if the list is empty) and click Connect. A progress window appears while the application performs a handshake with the main module. When it succeeds, the card’s status turns to connected.

  • If the handshake fails, check that the main module is in experiment mode and that no other program is using the port (Troubleshooting).

4. Choose the experiment folder

Click Select Folder (in Experiment Setup) and pick or create a folder for this session. The Start Experiment button stays disabled until you are connected and have chosen a folder. This is where the CSV files are written (Data output).

5. Start the experiment

Click Start Experiment. The button turns into Stop Experiment. The experiment runs as a sequence of phases shown under Current Phase, for example ITI, WAIT_TUNNEL, SHOW_STIM, WAIT_TOUCH, DELIVER_PELLET, WAIT_PELLET_RETRIEVAL.

In the Experiment Setup area you can see (and change before the start) the number of trials and the inter-trial interval. Stages from S2 have a stage setup area with the blink duration (how long the screens flash after an error) and the learning threshold.

6. Monitor

This is a stage S3 panel during a session (simulated values): connected card at the top, left and right performance, trial counters, sensors, actuators, animal information, phase and the log.

A stage S3 panel during a session, with simulated values

Each box panel shows:

Area

What it tells you

Summary card (top row)

Box name, connection status, current stage and phase, at a glance. Handy with several boxes.

Stage setup / Experiment Data

Trial counters and success rates for the stage: current trial (S1); total trials, correct responses, left and right touches and overall performance (S2); left and right performance, correct count and number of corrective loops (S3, S4).

Sensors

One indicator per sensor: nose-poke / pellet sensor, left and right response, left and right tunnel crossing, feeder alert. They light when the sensor is active.

Actuators

Reward, lights, and the left and right display controls. You can switch them by hand to test (for example Send Set Command). Avoid doing this during a real session.

Pellet alert (red text)

Appears if the application has had to retry pellet delivery several times: check the feeder.

Animal Information

The details you entered.

Stage / Phase info

Where the experiment is in the stage.

Log (bottom)

A running log, coloured by level (yellow information, orange warning, red error).

Several boxes at once

With --n-devices 2 (or more), one panel per box is shown, named Box 1, Box 2, …. The Broadcast Controls let you tick several boxes (Select All / Select None) and then Start Selected, Stop Selected or Set Folder for Selected in one click.

Two boxes with the Broadcast Controls bar

Stopping

Click Stop Experiment. The application stops the trial sequence, resets the outputs to a known state, logs Experiment ended, and closes the event log. Closing the window closes the serial connection.

What a stage looks like to you

  • The application waits for the animal, runs the trials, and updates the counters.

  • When the success rate reaches the learning threshold (see Training stages S1–S4), the stage ends and — if you chose All — the next one starts.

  • Every sensor change and every trial is recorded in the CSV files as it happens, and flushed to disk immediately, so a crash does not lose what was recorded before it.

Good practice

  • Do a short dry run without an animal after any change to the box or the software.

  • Keep a copy of the configuration file used for each cohort (Configuring experiments).

  • Record the software version and firmware version with your data.

  • Copy the CSV files to a safe location after each session and never edit the originals.